| Display | 7 inch colour TFT LCD, high resolution |
|---|
| Parameters | ECG / SpO2 / NIBP / RESP / TEMP |
|---|
| Patient types | Neonates to adult |
|---|
| ECG leads | Standard 3 leads RA, LA, LL; lead mode I, II, III. 5-lead cable RA, LA, LL, RL and V; lead mode I, II, III, aVR, aVL, aVF, V (7 leads) |
|---|
| ECG gain | 0.25, 0.5, 1, 2 |
|---|
| Heart rate range | 30–300 bpm |
|---|
| Heart rate accuracy | ±10% or 5 bpm, whichever is greater |
|---|
| Heart rate resolution | 1 bpm |
|---|
| ECG sensitivity | >200 µV |
|---|
| ECG input impedance | >5 MΩ |
|---|
| NIBP mode | Intelligent oscillation mode |
|---|
| Cuff inflating time | <30 sec |
|---|
| Initial cuff pressure | Adult 180 mmHg; neonate 120 mmHg |
|---|
| Maximum inflating pressure | 300 mmHg |
|---|
| Over-pressure protection | Adult 300 mmHg; paediatrics 240 mmHg; neonates 150 mmHg |
|---|
| NIBP accuracy | Maximum average error 4 mmHg |
|---|
| NIBP resolution | 1 mmHg |
|---|
| NIBP storage | 1000 NIBP measurement data storage |
|---|
| SpO2 sensor | Two-wave LED |
|---|
| SpO2 range | 0%–100% |
|---|
| SpO2 resolution | 1% |
|---|
| SpO2 accuracy | ±2% (70–100%); ±3% (50–69%) |
|---|
| Pulse rate range | 40–250 beat/min |
|---|
| Pulse rate accuracy | ±2 beat/min |
|---|
| Respiration method | Thoracic impedance |
|---|
| Respiration range | 10–100 rpm |
|---|
| Respiration accuracy | 1 bpm |
|---|
| Temperature range | 20–45°C |
|---|
| Temperature unit | °C/F selectable |
|---|
| Trend | 96 hours trend chart and alarm function |
|---|
| Compatibility | Can be used with ESU, defibrillator and pacemaker simultaneously |
|---|
| Optional | Recorder; upgradable to Masimo SpO2 |
|---|
| Battery | Built-in battery backup |
|---|
| Warning light | Big eyes warning light |