| Display | 12.1 inch TFT | 7 inch colour TFT LCD, high resolution |
| Parameters | SpO2, ECG, NIBP, Resp, Temp | ECG / SpO2 / NIBP / RESP / TEMP |
| Battery backup | Up to 2 hours | — |
| Made in India | Yes | — |
| Warranty | 2 years | — |
| Patient types | — | Neonates to adult |
| ECG leads | — | Standard 3 leads RA, LA, LL; lead mode I, II, III. 5-lead cable RA, LA, LL, RL and V; lead mode I, II, III, aVR, aVL, aVF, V (7 leads) |
| ECG gain | — | 0.25, 0.5, 1, 2 |
| Heart rate range | — | 30–300 bpm |
| Heart rate accuracy | — | ±10% or 5 bpm, whichever is greater |
| Heart rate resolution | — | 1 bpm |
| ECG sensitivity | — | >200 µV |
| ECG input impedance | — | >5 MΩ |
| NIBP mode | — | Intelligent oscillation mode |
| Cuff inflating time | — | <30 sec |
| Initial cuff pressure | — | Adult 180 mmHg; neonate 120 mmHg |
| Maximum inflating pressure | — | 300 mmHg |
| Over-pressure protection | — | Adult 300 mmHg; paediatrics 240 mmHg; neonates 150 mmHg |
| NIBP accuracy | — | Maximum average error 4 mmHg |
| NIBP resolution | — | 1 mmHg |
| NIBP storage | — | 1000 NIBP measurement data storage |
| SpO2 sensor | — | Two-wave LED |
| SpO2 range | — | 0%–100% |
| SpO2 resolution | — | 1% |
| SpO2 accuracy | — | ±2% (70–100%); ±3% (50–69%) |
| Pulse rate range | — | 40–250 beat/min |
| Pulse rate accuracy | — | ±2 beat/min |
| Respiration method | — | Thoracic impedance |
| Respiration range | — | 10–100 rpm |
| Respiration accuracy | — | 1 bpm |
| Temperature range | — | 20–45°C |
| Temperature unit | — | °C/F selectable |
| Trend | — | 96 hours trend chart and alarm function |
| Compatibility | — | Can be used with ESU, defibrillator and pacemaker simultaneously |
| Optional | — | Recorder; upgradable to Masimo SpO2 |
| Battery | — | Built-in battery backup |
| Warning light | — | Big eyes warning light |