| Display | 7 inch colour TFT LCD, high resolution | 10.1 inch WXGA touchscreen |
| Parameters | ECG / SpO2 / NIBP / RESP / TEMP | SpO2, 5-lead ECG, NIBP, Temp, Resp, EtCO2 optional |
| Patient types | Neonates to adult | — |
| ECG leads | Standard 3 leads RA, LA, LL; lead mode I, II, III. 5-lead cable RA, LA, LL, RL and V; lead mode I, II, III, aVR, aVL, aVF, V (7 leads) | — |
| ECG gain | 0.25, 0.5, 1, 2 | — |
| Heart rate range | 30–300 bpm | — |
| Heart rate accuracy | ±10% or 5 bpm, whichever is greater | — |
| Heart rate resolution | 1 bpm | — |
| ECG sensitivity | >200 µV | — |
| ECG input impedance | >5 MΩ | — |
| NIBP mode | Intelligent oscillation mode | — |
| Cuff inflating time | <30 sec | — |
| Initial cuff pressure | Adult 180 mmHg; neonate 120 mmHg | — |
| Maximum inflating pressure | 300 mmHg | — |
| Over-pressure protection | Adult 300 mmHg; paediatrics 240 mmHg; neonates 150 mmHg | — |
| NIBP accuracy | Maximum average error 4 mmHg | — |
| NIBP resolution | 1 mmHg | — |
| NIBP storage | 1000 NIBP measurement data storage | — |
| SpO2 sensor | Two-wave LED | — |
| SpO2 range | 0%–100% | — |
| SpO2 resolution | 1% | — |
| SpO2 accuracy | ±2% (70–100%); ±3% (50–69%) | — |
| Pulse rate range | 40–250 beat/min | — |
| Pulse rate accuracy | ±2 beat/min | — |
| Respiration method | Thoracic impedance | — |
| Respiration range | 10–100 rpm | — |
| Respiration accuracy | 1 bpm | — |
| Temperature range | 20–45°C | — |
| Temperature unit | °C/F selectable | — |
| Trend | 96 hours trend chart and alarm function | — |
| Compatibility | Can be used with ESU, defibrillator and pacemaker simultaneously | — |
| Optional | Recorder; upgradable to Masimo SpO2 | — |
| Battery | Built-in battery backup | Up to 2 hours |
| Warning light | Big eyes warning light | — |
| Network | — | IntelliVue LAN / Wi-Fi |
| Weight | — | 3.2 kg |