Nidek Bravo 10 Patient Monitor
Nidek

Nidek Bravo 10 Patient Monitor

Price on request

Nidek Bravo 10 is a 10.4-inch multi-parameter patient monitor for adults, paediatrics and neonates. Standard parameters are ECG, NIBP, SpO2, HR/PR, 2-RESP and 2-TEMP, with electrosurgical interference resistance and 360-hour trend review.

Specifications

Patient typesAdults, paediatrics and neonates
Display10.4" colour TFT LCD, high brightness, wide viewing
Parameters (standard)ECG, NIBP, SpO2, HR/PR, 2-RESP, 2-TEMP, rechargeable battery
OptionalThermal printer, Nellcor SpO2, 2-IBP, ETCO2 (mainstream or sidestream), wire or wireless network, wall mount, touch screen, 12-lead ECG, CO, anesthesia gas analysis
ESU interferenceSuper ability against electrosurgical interference
RespirationThoracic impedance & nasal tube double mode respiration test
Trend review360 hours history data
ECG heart rate range15–300 BPM
ECG accuracy1 BPM or ±1%
Heart rate alarmUpper 60–300 bpm (±10%); lower 10–200 bpm (±10%)
ECG frequency response (filtered)1–25 Hz
ECG frequency response (non-filtered)0.5–140 Hz
ECG sweep speed3.125, 6.25, 12.5, 25, 50 mm/s
ECG sensitivity selection2.5 (×¼), 5 (×½), 10 (×1), 20 (×2), 40 (×4) mm/mV, Auto
NIBP systolic40–250 mmHg (5.3–33.3 kPa)
NIBP diastolic10–210 mmHg (1.3–28 kPa)
NIBP mean20–230 mmHg (2.7–30.7 kPa)
NIBP accuracy≤±0.4 kPa or ±3 mmHg or ±2%
IBP range-60 to 500 mmHg
IBP channels2 channels
RR range0–156 BPM
RR accuracy±1 BPM or ±5%
TEMP channels2
TEMP range25–45°C
TEMP accuracy±0.2°C
TEMP response time≤2.5 min
SpO2 range0–100%
PR range15–300 BPM
PR accuracy±1 BPM or ±2%
CO2 parametersEtCO2, FiCO2, RR
CO2 range0–13% or 0–100 mmHg
CO2 resolution0.1% or 1 mmHg
Use casesPatient room, ER, OR, PACU, ICU, CCU