| Model | Luna Syringe Pump | TERRA |
| Loading type | Front loading | — |
| Syringe size | 5, 10, 20, 30, 50, 60 ml | — |
| Applicable syringe | Compatible with syringe of any standard | — |
| VTBI | 0.1–9999 ml | 1–9999 ml |
| Flow rate 5 ml | 0.1–100 ml/h (increments 0.01, 0.1, 1, 10 ml/h) | — |
| Flow rate 10 ml | 0.1–400 ml/h | — |
| Flow rate 20 ml | 0.1–600 ml/h | — |
| Flow rate 30 ml | 0.1–900 ml/h | — |
| Flow rate 50/60 ml | 0.1–1800 ml/h | — |
| Bolus rate 5 ml | 0.1–100 ml/h (0.1, 1 ml/h increments) | — |
| Bolus rate 10 ml | 0.1–400 ml/h | — |
| Bolus rate 20 ml | 0.1–600 ml/h | — |
| Bolus rate 30 ml | 0.1–900 ml/h | — |
| Bolus rate 50/60 ml | 0.1–1800 ml/h | — |
| Anti-bolus | Automatic | — |
| Accuracy | ±2% (mechanical accuracy <1%) | ±3% |
| Infusion modes | Easy Mode; Flow rate ml/min, ml/h; Time-based; Bodyweight | — |
| KVO rate | 0.1–2.00 ml/h (0.1 ml/h increments) | 0.1–5 ml/h (0.1 ml/h increments) |
| Occlusion sensitivity | High, medium, low | High, medium, low |
| Drug library | Available | — |
| Alarms | Occlusion, near empty, end program, not infusing, AC power off, motor malfunction, system malfunction, reminder alarm, pressure sensor error, syringe installation error, syringe drop off | Occlusion, air-in-line, door open, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm |
| Power supply | AC 110/230 V (optional), 50/60 Hz, 20 VA | — |
| Working temperature | 5–40°C | 10–40°C |
| Relative humidity | 20–90% | 30–75% |
| Atmospheric pressure | 700–1060 hPa | 700–1060 hPa |
| Size | 245 × 120 × 115 mm | 174 × 126 × 215 mm |
| Weight | 2.5 kg | 2.5 kg |
| Safety classification | Class II, type BF | Class I, type CF |
| Pumping mechanism | — | Curvilinear peristaltic |
| IV set | — | Compatible with IV sets of any standard |
| Flow rate | — | 0.1–1500 ml/h (0.1 ml/h increments) |
| Purge / bolus | — | 100–1500 ml/h (1 ml/h increments); purge when pump stops, bolus when pump starts |
| Bolus volume | — | 1–20 ml (1 ml increments) |
| Infusion mode | — | ml/h, drop/min, time-based |
| Air-in-line detection | — | Ultrasonic detector |
| Wireless management | — | Optional |
| Power supply AC | — | 110/230 V (optional), 50–60 Hz, 20 VA |
| Battery | — | 9.6±1.6 V, rechargeable |
| Battery life | — | 5 hours at 30 ml/h |