| Syringe compatibility | 1–60 mL (all major brands) | — |
| Flow rate | 0.1–1200 mL/h | 0.1–1500 ml/h (0.1 ml/h increments) |
| Accuracy | ±2% | ±3% |
| Alarms | Occlusion, syringe empty, low battery | Occlusion, air-in-line, door open, end program, low battery, end battery, AC power off, motor malfunction, system malfunction, reminder alarm |
| Drug library | 200+ drugs pre-loaded | — |
| Model | — | TERRA |
| Pumping mechanism | — | Curvilinear peristaltic |
| IV set | — | Compatible with IV sets of any standard |
| Purge / bolus | — | 100–1500 ml/h (1 ml/h increments); purge when pump stops, bolus when pump starts |
| Bolus volume | — | 1–20 ml (1 ml increments) |
| VTBI | — | 1–9999 ml |
| Infusion mode | — | ml/h, drop/min, time-based |
| KVO rate | — | 0.1–5 ml/h (0.1 ml/h increments) |
| Occlusion sensitivity | — | High, medium, low |
| Air-in-line detection | — | Ultrasonic detector |
| Wireless management | — | Optional |
| Power supply AC | — | 110/230 V (optional), 50–60 Hz, 20 VA |
| Battery | — | 9.6±1.6 V, rechargeable |
| Battery life | — | 5 hours at 30 ml/h |
| Working temperature | — | 10–40°C |
| Relative humidity | — | 30–75% |
| Atmospheric pressure | — | 700–1060 hPa |
| Size | — | 174 × 126 × 215 mm |
| Weight | — | 2.5 kg |
| Safety classification | — | Class I, type CF |